Technical Description

  The efficient control of the capacity and temperature in refrigeration systems is an important topic in refrigeration engineering. With ET 426 different methods of capacity control can be investigated. For this purpose a refrigeration circuit with two refrigeration chambers is available in which a cooling load is generated using an adjustable heater. Fans in both refrigeration chambers ensure an even temperature distribution.
  In the first refrigeration chamber a thermostat with solenoid valve and an evaporation pressure controller control the pressure of the refrigerant in the evaporator and thereby the temperature.
  Sensors record the temperature in the second refrigeration chamber and the temperature of the refrigerant upstream and downstream of the evaporator. A refrigeration controller processes the signals from the sensors. Via a solenoid valve the refrigeration controller changes the refrigerant flow in the evaporator and controls thereby the temperature in the second refrigeration chamber. The temperatures at which the fans and a defrost heater are switched on or off are also set at the refrigeration controller. In addition the refrigeration capacity of the overall refrigeration circuit can be adjusted via the speed of the compressor. There also is the option to adjust the refrigeration capacity at a capacity controller in the compressor bypass.
  Both refrigeration chambers feature a window to be able to monitor the fan operation and formation of ice. Manometers enable the easy reading of pressures at the relevant points.
  The well-structured instructional material sets out the fundamentals and provides a step-by-step guide through the experiments.

Learning Objectives / Experiments

- key devices for changing the refrigeration capacity
  * thermostat
  * evaporation pressure controller
  * refrigeration controller
  * capacity controller
  * compressor with variable speed
- operation of a refrigeration controller with

  specification of temperatures for
  * controlling the refrigeration capacity
  * switching the defrost heater on and off
  * switching the fans on and off

* Examination of different capacity control methods1
* Open compressor with variable speed1
* Adjustable cooling load via heater1
* Controlled defrosting with temperature specification
Alternative products
ET 165 -- Refrigeration System with Open Compressor
ET 412C -- Refrigeration System with Refrigeration and Freezing Chamber
ET 422 -- Capacity Control and Faults in Refrigeration Systems
ET 428 -- Energy Efficiency in Refrigeration Systems
[1] methods for capacity control in refrigeration systems
[2] refrigeration circuit with compressor, condenser and 2 evaporators in insulated refrigeration chambers
[3] each refrigeration chamber with adjustable heater to generate a cooling load, thermostatic expansion valve and fan
[4] 1 refrigeration chamber with refrigeration controller for temperature control; solenoid valve, fan and defrost heater as actuators
[5] 1 refrigeration chamber with thermostat, solenoid valve and evaporation pressure controller for temperature control
[6] compressor with variable speed via frequency converter
[7] capacity controller in the compressor bypass
[8] Refrigerant R134a, CFC-free
Technical Data

- refrigeration capacity: approx. 560W at -5/25C

  and 1.450min-1
- drive motor: 550W
2 heaters: approx. 500W
4 fans: max. air flow: approx. 50m/h
1 defrost heater: approx. 75W
Refrigeration controller
- 7 inputs
- 7 outputs
Evaporation pressure controller: 0...5,5bar
Thermostat: -5...20C
Capacity controller: 0,2...6bar


Measuring ranges
- motor speed: 1.100...2.700min-1

- temperature: 3x -60...50C
- heater power: 2x 0...1.000W
- pressure: 3x -1...9bar; 1x -1...24bar

Dimensions and Weight
LxWxH: 1.100x750x1.900mm
Weight: approx. 150kg
Required for Operation
230V, 50/60Hz, 1 phase
Scope of Delivery

1 trainer

1 set of instructional material

Order Details

061.42600  ET 426  Capacity Control in

                                 Refrigeration Systems

© 2005 G.U.N.T. Gerätebau GmbH