Products  |  Thermal Engineering and HVAC  |  Power Engines and Machines  |  ET 860
ET 860 Steam Boiler Safety & Control Show data sheet (pdf file) Print page
Technical Description

  The unit has been developed and built in accordance with the German technical regulations for steam and boilers (TRD 602).

  The normal operation of a boiler as well as malfunctions in the water level and pressure sections of the boiler that trip the safety control system can be demonstrated. In this context, the term "safety control system" is used to describe the following: the boiler is monitored and operated by safety devices connected in series. All parts used as safety device are original parts with TRD approval.

  In addition to the safety devices it is possible to switch in up to 15 system faults. Various fault conditions in system components can be simulated and their identification practiced.

  The measured values are transmitted directly to a PC via USB. The data acquisition software is included.

 

 

 
Learning Objectives / Experiments

- model of an oil-fired steam boiler with operating

  and safety components
- characteristic curves of the monitoring elements
- fault simulation
  * burner with flame monitoring
  * pressure switch and limiter
  * feed water and level controller
  * high and low water limiter

Features

* Complete boiler simulation with fault circuits and

  data acquisition1
* Transparent boiler, clear view of measuring and

  control devices

Alternative products
HL 392C -- Safety & Control in Heating Systems
Specification

[1] steam boiler simulation with oil burner
[2] simulation: standard boiler operation; 15 faults that trigger the safety control system
[3] safety control system according to TRD; safety elements: level switches, pressure switch and pressure limiter
[4] transparent boiler
[5] oil burner operated with compressed air
[6] front panel with process schematic, indicator lamps and lab jacks
[7] LabVIEW software for data acquisition via USB under Windows XP or Windows Vista

 

Technical Data

Burner

- light oil mechanical atomiser: 15kW
Boiler capacity: 110L
Supply tank: 150L
 

Pump

- power consumption: 40...70W

- max. flow rate: 66L/min

- max. head: 4m

Pressure switch: 0,5...6bar
Pressure limiter: 0,5...6bar
Safety valve: 6bar

 

Measuring ranges

- pressure: 0...6bar

- level: 0...100%

 

Dimensions and Weight
LxWxH: 1.820x790x1.800mm
Weight: approx. 150kg
Required for Operation
230V, 50/60Hz, 1 phase or 120V, 60Hz, 1 phase
Compressed air: 5bar
Scope of Delivery

1 trainer

1 LabVIEW software CD + USB cable

1 digital multimeter

1 set of laboratory cables

1 manual

Order Details
061.86000  ET 860  Steam Boiler Safety & Control